2025–2026 · Test automation

Thermal Controller & Logger

Building a Python workflow for multichannel trends and traceable test reports.

PROJECT BRIEF

My role and the result

Problem

Make repetitive temperature tests easier to monitor, compare, and document.

My contribution

Built the Python acquisition, trend-display, and reporting workflow; later documented recovered architecture and added a CSV audit utility.

Engineering tools

Python, Tkinter, Pico TC-08, thermocouples, CSV/Excel, and an optional Arduino serial interface.

Validation method

Inspected the recovered source and ran four passing synthetic checks of the separate CSV audit utility.

Documented result

Source supports eight thermocouple channels, live plots, test metadata, and CSV/Excel reporting.

Limitations

Hardware accuracy and acquisition timing were not revalidated. Displayed traces are synthetic; original Arduino firmware was not recovered.

Engineering decisions

Keep capture and reporting traceable

The application saves test metadata with channel readings and supports CSV plus optional formatted Excel output. Plotting retains a bounded history while logging continues independently, separating the display from the record of the test.

Make the trend rule explicit

The configurable indicator compares the maximum and minimum of a rolling window, using 10 valid readings and a 3.0 °C band by default. This makes the software rule inspectable, but it does not independently prove thermal equilibrium or sensor accuracy.

Testing and results

The recovered Python source documents an eight-channel measurement interface, trend plots, metadata capture, and CSV/Excel exports. Four synthetic tests passed for the later CSV audit utility. Those tests verify offline calculations, not thermocouple calibration or physical controller performance.

Next verification step

Record actual acquisition timestamps, validate the TC-08 and driver interface on hardware, and recover and test the optional Arduino control path.

THERMAL VALIDATION / SOFTWARE VISUALIZATION

A clearer view of the testing workflow.

The surviving application contains the acquisition, graphing, optional ambient-control interface, and reporting logic described below. These graphics were reconstructed from the publicly documented features. They are not captures of the original software running on physical equipment.

8 channelsExternal thermocouple inputs on the Pico TC-08
1.0 sNominal UI poll interval, not verified acquisition timing
10 readingsDefault rolling trend assessment window
3.0 °CDefault maximum-minus-minimum trend band

Instrument interface and live trending

The desktop application uses Tkinter for test configuration, channel selection and reporting. A separate graph window displays temperature trends with selectable channels. This illustration shows the type of interface, not a screenshot of the deployed application.

FIG 01 / APPLICATION
Conceptual dark-themed Thermal Logger interface with channel selectors and four colored illustrative temperature traces; all graph values are synthetic.
Source-informed interface illustration. Traces are synthetic examples, not recovered lab measurements, actual logger screenshots, or validated performance data. The original logger supports up to eight thermocouple channels and optional ambient-control functions.

How the data moves through the system

A Pico TC-08 supplies thermocouple readings to the Python application through the vendor Windows driver. The application coordinates graph updates, test metadata and exports. An optional serial connection can send setpoints and feedback to Arduino control hardware.

FIG 02 / SYSTEM PATH
Source-based system flow from up to eight thermocouples through Pico TC-08, Windows driver, Python Tkinter coordinator, and live plots or CSV and Excel export. An optional unverified Arduino controller branch is shown.
Functional architecture traced from public code and documentation. The optional Arduino firmware, wiring and heater behavior were not recovered or physically retested. A nominal one-second UI schedule is not proof of exact sample timestamps.
Data capture and reporting

Test metadata, selected temperature channels and sample numbers are exported to CSV, with an optional formatted Excel version.

Important timing limitation

Exported sample indices are not actual acquisition timestamps. Sampling jitter, driver accuracy and long-duration reliability require hardware testing.

How rolling stability is evaluated

The original display logic examines the range of the most recent valid readings. With the default settings, a 10-reading window is within the configured band when its maximum minus minimum is no greater than 3.0 °C.

FIG 03 / TREND RULE
Side-by-side simulated ten-sample thermal line graphs. Example A has a 1.2 degree Celsius range within a 3.0 degree configured band; Example B has a 5.3 degree range outside the band. Both plots are synthetic.
Mathematically calculated example windows using synthetic values. This figure explains a software display rule only; neither series is experimental data. The range check does not independently establish thermal equilibrium, sensor accuracy, calibration or repeatability.
Source-supported capabilities
  • Live multichannel trends and channel selection
  • Optional ambient setpoint interface
  • CSV metadata and optional Excel reports
Verification still needed
  • Pico TC-08 calibration and real sample timing
  • Arduino controller firmware and circuit
  • Original test-data provenance and hardware outcomes

Evidence & limitations

Evidence and publication boundary: The graphics describe recovered software structure and synthetic calculation examples, not actual thermocouple measurements. Internal engineering logs, employer records, proprietary reports and source files are not provided for download from this project page.