Capstone · 2026 · Health technology

IVerify

Leading systems integration for an external tubing-force sensing prototype.

PROJECT BRIEF

My role and the result

Problem

Explore restriction sensing without cutting or entering the sterile fluid pathway.

My contribution

Led systems integration for a four-person capstone team across sensing, analog electronics, microcontroller logic, and indication.

Engineering tools

Mechanical force sensing, signal conditioning, threshold logic, embedded supervision, and verification planning.

Validation method

Investigated sensor and threshold response in a non-clinical bench prototype with two LED indications.

Documented result

Demonstrated a basic two-LED indication path. The three-state interface remained a design goal.

Limitations

No clinical accuracy or safety validation. Later CAD and signal graphics are conceptual reconstructions, not the fabricated prototype.

Engineering decisions

Use an external mechanical interface

The clip-on concept applies preload outside the tubing and senses its reaction force. Keeping the sensor outside the fluid pathway was a defining design constraint; translating that force into reliable restriction detection still needed verification.

Coordinate the entire indication path

The integration work connected mechanical contact, analog conditioning, threshold response, and microcontroller supervision. The documented bench outcome was two LED indications; normal, partial-restriction, and occlusion states remained the intended product interface.

Testing and results

The student bench prototype demonstrated a basic path from sensing and threshold response to two indicator LEDs. It established a limited integration proof of concept; it did not establish a clinically reliable three-state occlusion detector.

Next verification step

Characterize preload repeatability, calibrate the force response, and test candidate thresholds across tubing and operating conditions before making any accuracy or clinical-use claim.

PUBLIC ENGINEERING WORK / CONCEPT RECONSTRUCTION

Inside the IVerify design concept

The student prototype and the later concept reconstruction represent different stages of work. The illustrated mechanical layout, sensing path and synthetic trend explain the proposed system without implying that the depicted hardware was built, calibrated or clinically tested.

Dark-pink and cyan exploded concept illustration of an IVerify clamshell, IV tubing, contact plunger and placeholder indicator electronics. The mechanical reconstruction is not original manufactured CAD.
Fig. 1. Visual reconstruction of the proposed clamshell concept. Geometry and component placement are illustrative; this is not the original team design, a fabricated assembly photograph, or validated medical equipment.
MECHANICAL CONCEPT / EDUCATIONAL

Exploded assembly concept

The illustration shows how external tubing, two jaw pieces, a contact plunger and placeholder electronics could fit together. This is a communication model for the concept, not proof of a functional clamp or sensor.

  • 3.0 mm tubing used as an assumed reference size.
  • 46 × 30 × 18 mm combined jaw envelope, invented for the reconstruction.
  • Later concept geometry contains nine modeled solids; their placement is not verified against a fabricated prototype.
  • No proven spring preload, fabricated latch, calibrated sensor, or tested tolerance stack-up.

Engineering targets, not verified performance

These requirements and ranges are documented in the publicly available reconstruction notes as original design goals. They are not presented as completed product qualification or clinical detection results.

5 NNominal mechanical preload target
3–8 NIntended adjustable preload range
0–20 NTarget force-sensor measurement range
≥50 HzTarget sensing/sample frequency

How the proposed sensing path works

This is the functional architecture of the student concept, not a reproduced circuit diagram or a clinically established relationship between restriction and sensed force.

  1. 01 · CONTACTPreloaded tubingMechanical reaction at the clamp
  2. 02 · SENSEForce sensorExternal force into an electrical signal
  3. 03 · CONDITIONAnalog front endGain and filtering
  4. 04 · DECIDEThreshold + MCUCandidate events and timing logic
  5. 05 · INDICATELED statusTwo LEDs on the reported bench prototype

What is demonstrated versus proposed

Spring 2026 student prototype

Bench-level investigation of external tubing-force sensing, analog threshold response, microcontroller supervision, and two-LED indications. These results do not establish clinical accuracy or safety.

Later reconstructed model

Parametric, AI-assisted concept geometry informed by the public project description. The removable jaw and three-LED board are illustrative design changes, not recovered original CAD or proven hardware.

Evidence & limitations

Evidence boundary: No quantitative clinical sensitivity, false-alarm rate, validated flow classification, or approved patient-use capability is claimed. The original team documents are not distributed through this website. Educational engineering concept only; not for IV therapy or clinical decisions.